作 者:张晗 赵真臻 张正清[1] 张琦[1] 甘桂芳 ZHANG Han;ZHAO Zhenzhen;ZHANG Zhengqing;ZHANG Qi;GAN Guifang(Department of Clinical Laboratory,Shanghai Ninth People′s Hospital,Shanghai Jiao Tong University School of Medicine,Shanghai 200011,China)
机构地区:[1]上海交通大学医学院附属第九人民医院检验科,上海200011
出 处:《国际检验医学杂志》2023年第S02期91-96,共6页International Journal of Laboratory Medicine
基 金:上海市“医苑新星”青年医学人才培养资助计划(沪卫人事[2020]087号)。
摘 要:目的对脂肪酶(LPS)和胆碱酯酶(CHE)试剂盒在B品牌全自动生化分析仪上的检测性能进行分析评价,确保检测结果的准确性。方法参照CNAS和WS/T等相关系列文件设计性能验证方案,对B品牌测定脂肪酶和胆碱酯酶的精密度、正确度、线性区间、可报告范围和生物参考区间进行验证及评价。结果LPS的低、高水平质控品的批内不精密度变异系数(CVr)和总不精密度的变异系数(CV1)值分别为:4.32%、4.10%,1.64%、1.73%;CHE的低、高水平质控品的CVr和CV1分别为:1.31%、1.27%,0.98%、0.88%;检测实验室能力验证(PT)样本的LPS和CHE水平验证正确度,偏倚均符合要求;LPS线性区间在4.63~512.60 U/L内良好(Y=0.9945 X+0.4739,r 2=0.9997)、最大可报告范围扩展至1537.80 U/L,CHE线性区间在1.27~14.75 kU/L内良好(Y=1.004 X-0.2436,r 2=0.9978)、最大可报告范围扩展至29.50 kU/L;试剂说明书提供的生物参考区间均验证通过(符合率均≥95%)。结论B品牌全自动生化分析仪检测LPS和CHE的精密度、正确度、线性区间、临床可报告范围和参考区间性能验证良好,能够规范检测、确保检测质量,满足临床需求。Objective To analyze and evaluate the detection performance of lipase(LPS)and cholinesterase(CHE)test kits on the B brand fully automatic biochemical analyzer,in order to ensure the accuracy of the detection results.Methods The performance verification scheme was designed according to CNAS and WS/T documents.The precision,trueness,linear interval,clinical reportable range,and biological reference interval of LPS and CHE were tested and evaluated by Backman AU5800 automatic biochemical analyzer.Results The Intra-batch unrefined density(CVr)and the total unrefined density(CV1)of low and high quality controls for LPS were 4.32%,4.10%,1.64%,1.73%,respectively;The CVr and CV1 of low and high quality controls for CHE were 1.31%,1.27%,0.98%,0.88%,respectively.To verify the trueness,the levels of LPS and CHE in Proficiency Testing(PT)samples were detected,and the biases can meet the requirements.The linear interval of LPS was good within 4.63-512.60 U/L(Y=0.9945 X+0.4739,r 2=0.9997),and the clinical reportable range can be extended to 1537.80 U/L.The linear interval of CHE was good within 1.27-14.75 kU/L(Y=1.004 X-0.2436,r 2=0.9978),and the clinical reportable range can be extended to 29.50 kU/L.The biological reference intervals of LPS and CHE provided by the reagent manuals were verified(coincidence rate≥95%).Conclusion The precision,accuracy,linear interval,clinical reportable range,and reference interval performance of the B brand fully automatic biochemical analyzer for detecting LPS and CHE have been well validated,which can standardize the detection,ensure the quality of the detection,and meet clinical needs.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:18.188.195.92