数字减影血管造影球管工作原理及质量维护  

Digital subtraction angiography bulb tube working principle and quality maintenance

在线阅读下载全文

作  者:沈彧枫 曹少平 SHEN Yufeng;CAO Shaoping(Department of Medical Equipment,Yueyang Hospital Affiliated to Shanghai University of Traditional Chinese Medicine,Shanghai 200437,China)

机构地区:[1]上海中医药大学附属岳阳医院医学装备处,上海200437

出  处:《外科研究与新技术(中英文)》2024年第3期249-253,共5页Surgical Research and New Technique

基  金:上海市临床重点专科2019年评审项目(shslczdzk04601);上海交通大学中国医院发展研究院医院后勤管理研究所2024年度项目(CHDI-2024-C-14)。

摘  要:基于数字减影血管造影(DSA)球管工作原理,通过分析工作实践中的3个典型DSA球管故障,查找出故障主要成因,并概括了DSA球管的常见故障类型;总结出在安装新球管时的注意事项,如需仔细检查油泵接线,并对球管进行必要的训练和预热,以确保其稳定运行;最后详细描述了DSA球管各部位的维护保养及验收流程和标准。After the introduction of the working principle of digital subtraction angiography(DSA)bulb tube,through the analysis of three typical faults of DSA bulb tube,the main causes of the faults were found,and the common fault types of DSA bulb tube were summarized.It was concluded that when installing a new bulb tube,it was necessary to check the wiring of the oil pump carefully,and carry out the necessary training and preheating of the bulb tube to ensure its stable operation.Finally the maintenance and acceptance procedures and standards of the various parts of DSA bulb tube were described in detail.

关 键 词:数字减影血管造影 球管故障 预防性维护 灯丝开路 验收标准 

分 类 号:R02[医药卫生]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象