作 者:Robert Edward T.Ang Edward Kenneth F.Reyes Fernando Amado J.Ayuyao Jr Maria Isabel N.Umali Emerson M.Cruz
机构地区:[1]Asian Eye Institute,Phinma Plaza,Rockwell Center,8Th Floor,1200 Makati,Philippines [2]Cardinal Santos Medical Center,10 Wilson St.,Greenhills,1502 San Juan,Philippines
出 处:《Eye and Vision》2024年第5期28-40,共13页眼视光学杂志(英文)
摘 要:Background:To compare the measurements obtained from the Orbscan II,IOLMaster 700,Pentacam AXL,and Castroviejo caliper and their effects on calculating the recommended implantable collamer lens(ICL)size and postoperative vault measurements.Methods:This is a retrospective cross-sectional study of patients who underwent ICL surgery by a single surgeon from March 1,2018 to July 31,2021.Records were reviewed for the anterior chamber depth(ACD)and white-to-white(WTW)measurements obtained from the Orbscan II,IOLMaster 700,Pentacam AXL,and Castroviejo caliper(WTW only).These were used to calculate the recommended ICL size.The actual ICL size implanted,and vault measurements obtained one month postoperatively were also collected.Results:One hundred seven eyes with a mean age of 27.9±7.7 years were included in the study.Mean WTW measurements were significantly different between devices(P<0.0001),with the IOLMaster 700 having the highest value(12.14±0.04 mm)and the caliper having the lowest value(11.45±0.04 mm).Mean ACD measurements were the lowest in Orbscan II(3.12±0.25 mm)and the highest in Pentacam AXL(3.16±0.24 mm).The Pentacam AXL produced an ICL size similar to the Orbscan in 69.2%of eyes.The IOLMaster yielded an ICL measurement one size larger than Orbscan-based calculations in 64.5%of eyes.Using the Orbscan WTW and ACD,the desired vault of 0.25 to 0.75 mm and 0.25 to 1.00 mm was achieved in 70%and 91%of eyes,respectively.Substituting caliper WTW to IOLMaster 700 or Pentacam AXL WTW increases the percentage of achieving the desired vault to 80%,similar to the Orbscan.Conclusions:The Orbscan II,IOLMaster 700,and Pentacam AXL cannot be used interchangeably for calculating ICL sizing.Combining the WTW from caliper measurement with the ACD of the IOLMaster 700 or Pentacam AXL could improve ICL sizing and achieve a higher percentage of eyes with the desired vault.
关 键 词:ICL vault Orbscan II IOLMaster 700 Pentacam AXL CALIPER
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.49