离子色谱法测定人工肾透析液中阳离子和阴离子  被引量:1

Determination of Cations and Anions in Dialysis Solution for Artificial Kidney by lon Chromatography

在线阅读下载全文

作  者:翟所迪[1] 欧阳津[2] 关翠林[2] 张现化[2] 

机构地区:[1]北京大学第三医院 [2]北京师范大学

出  处:《生命科学仪器》2003年第5期30-34,共5页Life Science Instruments

摘  要:目的:建立一种快速测定人工肾透析液中的阳离子Na^+、K^+、Ca^(2+)、Mg^(2+),和阴离子Ac^-、Cl^-的新方法。方法:分别采用单柱阳离子色谱法和双柱阴离子色谱法测定人工肾透析液中的阳离子和阴离子。结果Na^+、K^+、Ca^(2+)、Mg^(2+)、Ac^-和Cl^-的线性范围分别是0.01~10,0.1~10,0.05~10,0.05~10,0.09~20,0.03~10ug·mL^(-1),相关系数r分别为0.999 8,0.9996,0.007 8,0.996 4,0.9995,0.999 6(n=5),每种离子测定结果的RSD均小于4%。结论:此方法简便快速,灵敏准确,优于其它方法。Objective:To introduce a new rapid detennination method of cations and anions in dialysis solution for artificial kidney.Method.The single column cationioc coromatography and the double columns anionic chronatograny were used to determine cations(Na+,K+,Ca2+,Mg2+)and anions (Ac and CI)respectively,Results Linear responses have been obtained in the following ranges for Na+,K+,Ca2+,Mg2,Ac and CI,respectively:0.01-10(t=0.9996),0.05-10(r=0.9978).0.05-10(r=0.9964).0.09-20(r=0.9995),0.03-10μg·mL-1(r=0.9996).The reprodueibility was also good(RSD<4%.n=5).Conclusion.The results showed that it is a quicl and simple method with convenience,accuracy and sensitivity,Therefore,it can be applied in rapid qualitative and quantitative determination of cations and anions in dialysis solution for artilicial kidney.

关 键 词:人工肾透析液 阳离子 阴离子 测定 离子色谱法 

分 类 号:R692[医药卫生—泌尿科学] O657.75[医药卫生—外科学]

 