人工晶体眼A常数的计算和修正  被引量:1

Calculation and Correction of the Constant A of Pseudophakia

在线阅读下载全文

作  者:徐雯[1] 姚克[1] 陈佩卿[1] 

机构地区:[1]浙江医科大学附属第二医院眼科

出  处:《浙江大学学报(医学版)》1999年第3期107-109,共3页Journal of Zhejiang University(Medical Sciences)

摘  要:目的:探讨适合国内患者的人工晶体屈光度计算的A常数值。方法:统计分析了355只后房型人工晶体眼的生理参数,并用二元线性回归对355只后房型人工晶体眼进行A常数计算。结果:康明、Storz和Pharmacia人工晶体的A常数分别为116.9、117.6和117.8,回归标准分别为2.44、1.83和2.98。结论:利用国内患者的临床资料进行二元线性回归而得的A常数值可指导临床医师更准确地计算人工晶体屈光度值。Objective: To probe into the constant A which is suitable for the calculation and correction of it among internal patients with pseudophakia. Methods: Statistical analysis of physiological parameters obtained from 355 patients with pseudophakia was made and the constant A among 355 patients with pseudophakia was calculated with dual linear regression. Results: The constants A of KM,Storz and Pharmacia were 116.9,117.6 and 117.8 respectively while standard deviation was 2.44,1.83 and 2.98 respectively. Conclusion: The constant A obtained by exploiting clinic data gathered from internal patients can be helpful to those clinic doctors in calculating the diopter of IOL more accurately.

关 键 词:人工晶体眼 A常数 计算 修正 屈光度 人工晶体植入术 白内障 

分 类 号:R776.1[医药卫生—眼科]

 