Validation of a conductometric bienzyme biosensor for the detection of proteins as marker of organic matter in river samples  

Validation of a conductometric bienzyme biosensor for the detection of proteins as marker of organic matter in river samples

在线阅读下载全文

作  者:KHADRO Basma NAMOUR Philippe BESSUEILLE Franois LEONARD Didier JAFFREZIC-RENAULT Nicole 

机构地区:[1]University of Lyon, Laboratory of Analytical,CNRS UMR 5180, 69622 Villeurbanne, France [2]Cemagref of Lyon

出  处:《Journal of Environmental Sciences》2009年第4期545-551,共7页环境科学学报(英文版)

基  金:supported by CEMAGREF Laboratory of Lyon

摘  要:This article describes a conductometric bi-layer based bienzyme biosensor for the detection of proteins as a marker of organic matter in rivers. Proteins were chosen to be used as indicators of urban pollution. The working mechanism of the bienzyme biosensor is based on the enzymatic hydrolysis of proteins into several fractions (peptides and amino acids), which results in a local conductivity change depending of the concentration of proteins. In this work, we began with the optimization of biosensor response using bovine serum albumin (BSA) as standard protein. For this objective seven enzymatic biosensors were prepared: four enzymatic sensors with only one layer of enzyme (proteinase K, trypsin, pronase or protease X) and three other enzymatic sensors with two layers (first layer: membrane containing proteinase K; second layer: one of the three other enzymes: trypsin, pronase or protease X). The biosensors were obtained through the deposition of enzymatic layers and the cross-linking process between enzymes and BSA in saturated glutaraldehyde vapour. The response of the various biosensors, described previously, were compared with the values of total organic carbon (TOC), and those of organic nitrogen (No~), as determined by the laboratory accredits (CEMAGREF of Lyon) using the traditional method of analysis (NF EN 1484, infrared spectroscopy) and (NF EN 25663, mineralization/colorimetry assay) respectively for each water sample obtained from different sites in Lyon (France). The linear correlations obtained with the response of the seven biosensors showed the most important indices of correlations for the biosensor with two enzymatic layers: proteinase K + pronase (pkp). The optimum conditions for the preparation of the pkp biosensor increased the sensitivity and gave a limit of quantification of 0.583 μg/L for TOC and 0.218 μg/L for Norg in water samples. This sensor shows good reproducibility (2.28%), a capacity to be used at temperatures range 10-This article describes a conductometric bi-layer based bienzyme biosensor for the detection of proteins as a marker of organic matter in rivers. Proteins were chosen to be used as indicators of urban pollution. The working mechanism of the bienzyme biosensor is based on the enzymatic hydrolysis of proteins into several fractions (peptides and amino acids), which results in a local conductivity change depending of the concentration of proteins. In this work, we began with the optimization of biosensor response using bovine serum albumin (BSA) as standard protein. For this objective seven enzymatic biosensors were prepared: four enzymatic sensors with only one layer of enzyme (proteinase K, trypsin, pronase or protease X) and three other enzymatic sensors with two layers (first layer: membrane containing proteinase K; second layer: one of the three other enzymes: trypsin, pronase or protease X). The biosensors were obtained through the deposition of enzymatic layers and the cross-linking process between enzymes and BSA in saturated glutaraldehyde vapour. The response of the various biosensors, described previously, were compared with the values of total organic carbon (TOC), and those of organic nitrogen (No~), as determined by the laboratory accredits (CEMAGREF of Lyon) using the traditional method of analysis (NF EN 1484, infrared spectroscopy) and (NF EN 25663, mineralization/colorimetry assay) respectively for each water sample obtained from different sites in Lyon (France). The linear correlations obtained with the response of the seven biosensors showed the most important indices of correlations for the biosensor with two enzymatic layers: proteinase K + pronase (pkp). The optimum conditions for the preparation of the pkp biosensor increased the sensitivity and gave a limit of quantification of 0.583 μg/L for TOC and 0.218 μg/L for Norg in water samples. This sensor shows good reproducibility (2.28%), a capacity to be used at temperatures range 10-

关 键 词:conductometric biosensor PROTEINS ENZYME natural water analyses total organic carbon organic nitrogen 

分 类 号:X832[环境科学与工程—环境工程]

 