动态导航技术中植入位点骨质量对种植精度的影响  被引量:8

The effect of bone quality of implant sites on implant accuracy in dynamic navigation technology

在线阅读下载全文

作  者:王磊[1] 伍颖颖[1] 满毅[1] WANG Lei;WU Yingying;MAN Yi(State Key Laboratory of Oral Diseases&National Clinical Research Center for Oral Disease,Department of Oral Implantology,West China Hospital of Stomatology,Sichuan University,Chengdu 610041,Sichuan Province,China)

机构地区:[1]口腔疾病研究国家重点实验室,国家口腔疾病临床医学中心,四川大学华西口腔医院口腔种植科,四川成都610041

出  处:《口腔颌面外科杂志》2022年第3期174-181,共8页Journal of Oral and Maxillofacial Surgery

基  金:国家自然科学基金(81970966);四川大学华西口腔医院新技术项目(LCXJS2018-6)。

摘  要:目的:探究颌骨骨密度和种植位点骨壁斜坡角度对动态导航技术中种植精度的影响。方法:对2017—2021年通过在上前牙及上颌前磨牙区采用动态导航技术种植的患者进行回顾分析,共纳入42例患者及47颗种植体。测量术区颌骨密度及种植位点骨壁和种植体长轴形成的角度,并计算术前设计和术后种植体位置的偏差。结果:统计结果显示,骨壁斜坡与种植体长轴的交角和种植体轴向的角度偏差具有显著的负相关关系(r=-0.298,P<0.05),种植术区颌骨密度与种植体轴向的角度偏差具有显著的负相关关系(r=-0.289,P<0.05)。种植体颈部骨宽度和种植体精度之间不存在相关关系(P>0.05)。结论:种植体长轴与颌骨骨壁形成的角度越小、种植术区颌骨密度越低,动态导航技术种植精度中的种植体轴向偏差越大。Objective:To explore the effects of bone density and angle of bone wall of the implant site on the implant accuracy under the dynamic navigation technology.Methods:A retrospective analysis was conducted in patients who underwent implantation with dynamic navigation technology in the upper anterior tooth and premolar regions from 2017 to 2021.A total of 42 patients and 47 implants were included.Measuring the bone density of the implant region and the angle between the bone wall of the implant site and the long axis of the implant,and calculating the deviation of the preoperative design and postoperative implant position.Results:Statistical results showed that there was a significant negative correlation between the slope of the bone wall and the intersection angle of the long axis of the implant and the axial angle deviation of the implant(r=-0.298,P<0.05).There was a significant negative correlation between the regional jaw bone density in the implant region and the axial angle deviation of the implant(r=-0.289,P<0.05).There was no correlation between implant neck bone width and implant accuracy(P>0.05).Conclusion:The results showed that the smaller the angle formed between the long axis of the implant and the mandibular bone wall,and the lower the mandibular bone density in the implant region,the greater axial deviation of the implant accuracy in the dynamic navigation technology.

关 键 词:动态导航 牙种植 骨密度 精确度 误差分析 

分 类 号:R782[医药卫生—口腔医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象